To the Kawaguchi electric works Ltd. in case of the thunder harm product (security machine, arrester, and SPD)

 
Communications
Equipment
  Protectors for Communication
  Search by Specifications
  Test Springs
  Terminal Blocks
  Installation Tools
  Protectors for Power Supply
Measuring Instrument
  Electrostatic Paper Analyzer: EPA8300A
  High Level Surface Potential Meter: S2002A
  Universal Electrometer: MMA II-17B
  Instruction of Examples of Use
 
Home > List of Products > Measuring Instruments > Introduction of examples of use > Induction electrode SP-04 Example of measurement of surface potential of specimen
 

Measuring Instruments

 

Induction electrode SP-04

Example of measurement of surface potential of specimen


Products → MMAII17B
 

DC high-voltage power supply V7032A for calibration Connection of 1 kV to the metal plate sets the reference voltage.

When MMA II-17B indicates 1V, calibration is conducted by 1 kV.
Then V7032A is set at 0 V, and the specimen is placed in the metal plate.
When MMA II-17B indicates 10 V, the surface potential of the specimen is 10 kV.